SOP: Preparation of Uncut Sheet for Lateral Flow Rapid Test
1. Materials Required
- Nitrocellulose membrane: Core of detection zone
- Sample pad: Receives the sample
- Conjugate pad: Holds labeled conjugates (e.g., gold-labeled antibodies)
- Absorbent pad: Draws the liquid through
- Backing card (adhesive plastic film): Provides support
- Capture antibodies/antigens: For Test line (T-line)
- Control line antibodies (e.g., anti-IgG): For Control line (C-line)
- Conjugate (e.g., gold nanoparticles conjugated to antibodies): For signal detection
- Buffers, stabilizers, surfactants: For treatment and storage
2. Equipment Needed
- Dispenser/striping machine (for T-line/C-line)
- Conjugate spraying machine
- Laminating/assembly machine
- Cutter (optional if not cutting into strips)
- Incubator or drying oven (temperature-controlled)
- Desiccator or dry storage
3. Procedure
Step 1: Membrane Striping (Test & Control Lines)
- Stripe test and control lines on nitrocellulose membrane using micro-dispensing system.
- Dry at 37-45°C for 30-60 minutes.
Step 2: Conjugate Pad Preparation
- Impregnate with gold-labeled conjugate.
- Dry at 37°C for 2-4 hours.
Step 3: Pad Treatment
- Treat sample and absorbent pads with buffer solution.
- Dry thoroughly.
Step 4: Assembly
- Layer in order: Sample pad -> Conjugate pad -> NC membrane (with T/C lines) -> Absorbent pad.
- Ensure slight overlaps for capillary flow.
Step 5: Final Drying & Storage
- Dry entire uncut sheet at 37°C for 1-2 hours.
- Store in desiccator or sealed pouch with desiccant.
★Tips
- Conduct in cleanroom environment (Class 10,000 or better).
- Maintain low humidity (<30% RH) during drying and storage.
- Perform quality control on each batch (flow time, line intensity, etc.).